CHARACTERISTICS
2
-g
CHARACTERISTICS
SPECIFICATIONS
2.12
CURSORS (MODELS
95 AND
97 ONLY)
Horizontal
Display Resolution
Digital
Readout Resolution
Error Limit
Cursor
Range
Vertical
Display Resolution
Digital Readout Resolution
Error Limit
25
parts
per div
3 digits
±0.1%±1
LSB
Visible
part of signal
25 parts per dIv
3 digits
±2%
ADDITIONAL INFORMATION
Cursors
cannot pass each
other.
Referred to Input
at BNC or Probe
tip, after Probe recalibration.
2.13 MULTIMETER
The Multimeter
uses the Channel A input
for VDC & VAC measuremenlsand
the
Safety Banana Jack
Inputs for Resistance,
Diode Test,
Continuity, and DC mV measurements.
An Internal
reference is
used to optimize the
accuracy of the Channel
A Input and any
probes used. The accuracy
of
all
Multimeter
measurements
Is within
± {(%
of reading)
+
(number of least-significant
digits)) from
18 *C
to 28
®C
with relative
humidity up to 20% fora period
of one year after calibration.
Add 0.1 x
{specified
accuracy)/C
for 18 or 28
”C Ambient.
•
Displayed
range include used
probe, if calibrated.
•
Values listed
are without
attenuating probe.
•
A Vrms
AC and V DC
dual display mode Is optimized
for power line
(mains) related
measurements.
•
DC
Voltage
Ranges
300 mV. 3V. 30V
& 300V Manual
or automatic ranging
on
peak voltage.
High Voltage x10
Probe extends
measurement
to
600V. Peak
voltage is 2.5x range,
except
375V In 300V range.
Resolution
0.1 mV, 1 mV,
0.01V,
&0.1V
Multiply
x10 with High Voltage
Probe-
Accuracy
±
(0.5% +
5)
Digital
Display
3000
counts
Up to 4500 counts,
3500 counts in
300V range.
Display update
Response
Time
<
300 ms
<3.5s
Zeroing
automatic
Series Mode
Rejection
Ratio
> 50
dB @ 50 Kz
or 60 Hz