EasyManua.ls Logo

Fluke 93 - External

Fluke 93
267 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
18
CIRCUIT DESCRIPTIONS
3.4.4 EXTERNAL (BANANA) INPUT/OUTPUT circuitry
Introduction
See figure 3.10.
The ScopeMeter Is provided with two Panana connectors, which are used
as
Inputs In
the mV,
DIODE, and OHM METER modes or as EXTerna I trigger inpirt in SCOPE mode. These connectors
also serve
as outputs for the built-in
generator.
Protection circuitry is provided to prevent damage
by
overvoltage.
t<27E1a
*0
Text, mv
Figure 3.10 Schematic diagram signal flow fn EXTemaf (banana) input/output circuitry
Detailed circuit deacription
See figure 3.1
0
and circuit diagram A2b (figure
10.6).
mV DC measurement circuitry
The
mV DC input
voltage
on the red banana terminal is
fed
to die L.F. pan of the channel A attenuator
section, via the followng path: R2750, K2750a, K2751 b, R2761
,
D2751 (refer to circurt diagram A2b,
figure 10.6). M/hen the ScopeMeter is switched to mV DC measurement using the EXT banana
terminals, the settings are
as
follows:
Table $.4 A-ASIC and attenuator settings in
mV
DC
mode.
mV
DC
RANGE A-ASIC (02301) LF-ATTENUATOR (channel A)
300
mV
I
100
mV/div
1-
3 V 100 mV/dIv O.V

Table of Contents

Related product manuals