4-4
PERFORMANCE VERIFICATION PROCEDURE
4. Vertical deflection coefficients channei A
These tests check the vertical deflection coefficients for cttannel A in the
100 mV/div DC and
AC
ranges.
Test ec^uipment:
Fluke 5100B Calibrator
Test setup:
Procedure/requlremente for AC teat:
A Apply
a
1 kHz sine wave signal with an amplitude of 600 mV AC peak-to-peak to the channel A
BNC connector.
(Set the Fluke 5100B to 212.13 mV RMS, 1 kHz sine wave).
Verify that the amplitude of the sine wave signal dieolayed is 5.68...6.12 divisions.
Procedure/requirements DC test:
B
Apply
300
mV DC to channel A.
Verify that the distance between the trace for channel A and the vertical middle of the screen
(ground level) Id 2.94.. .3.06 divisions.
5/6/7. Vertical deflection coefficients ctiannel D
These tests check the vertical deflection coefficients for channel 6 In the DC and
AC
ranges.
Test equipment:
Fluke
5100B Calibrator