EasyManua.ls Logo

Fluke 93 - Page 80

Fluke 93
267 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PERFORMANCE VERIFICATION PROCEDURE 4-
11
Test set-up:
Procedure:
A Apply
a
2 kHz sine
wave signal of 800 mV peak-to-peaK to channel A and channel B.
Use 50^
terminations.
Adjust the Input signal to
a trace height of 8 divisions.
Requirements:
Verify that
a
line with
an angle of
45°
is displayed.
See
figure 4.8.
Figure 4.8
A versus 8
display
21/22. Base line instability
This test checks the maximum
base line
instability.
Teat equipment:
none
Test setup:
no special
setup
required