EasyManua.ls Logo

Fluke 93 - Page 89

Fluke 93
267 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-20
PERFORMANCE VERIFICATION PROCEDURE
4. Power supply voltage range
•**
Ail models
*'•
Th^ test checks the correct operation of the
ScopeMeter within the boundaries of the DC supply
voltage.
Test equipment:
Philips PE 1537 Power Supply 0-40V/0-1A
Tektronix SO 503 Constant Ampittude Sine Wave Generator
5 mm Power Jack connector plug with attached cable {for example order 4622
321 20125)
Test
set-up:
ScoQ«Uetar
Settl ngs/proeedure
:
A Insert the power
plug into the power adapter contact on the side of the ScopeMeter.
B Switch on the power
supply and set the voltage to a wanted value between 8 and 20V DC.
C
Apply a 50 kHz
sine
wave v/tth an amplitude of 100 mV peak-to- peak to channel
A.
Use
a SOD
termination.
D Swich on the ScopeMeter. At power on, a beep tone must
be
audible.
E Press AUTO SET and verify that a well triggered signal with an amplitude of approximately
5
divisions is displayed over the whole supply voltage range.
Requirements:
A The ScopeMeter must start
at any DC voltage between 3 and 20V, applied at Its power adapter
contact.
B The ScopeMeter must remain operative
over the indicated voltage range.
C The amplitude of the trace displayed
must be approximately 5 divisions, independent of the
supply voltage.
Figure 4.10 Power Jack connector

Table of Contents

Related product manuals