4-24
PERFORMANCE VERIFICATION PROCEDURE
Test setup:
no specific set*
up
required
Setti ng/procedure
:
A Switch on the ScopeMeter and switch
to SCOPE mode.
Operate the keys to get
a
from setting
that differs frorr the default settings:
Set channel A and B la
500
mV/di
v.
Set the timebase to 1 ms/div.
B Press the SETUP key to gel into the SETUP
menu
C
Press the
SAVE softkey, select SETUP 3 from the pop-up menu, and press ENTER. This will
save the current front setting as SETUP 3.
D Set channel A and B lo 2V/div. Set the timebase to
1
^e/div.
E
Switch off the ScopeMeter.
F Switch on the ScopeMeter again (do not use MASTER RESET!).
Press the SETUP Key to get
into the SETUP menu.
G Press the RECALL softkey and choose SETUP
3
from
the pop-up menu. (Use the select/adjust
keys and the ENTER softkey.) This entry
is marked in the pop-up menu. The frortt setting must
be restored to the setting previously selected
In step A.
H Now press the
DELETE softkey. Use the select/adjust key and the ENTER softkey to choose
SETUPS from the pop-up menu. The RECALL marker svill disappear now
as a
sign
that the front
setting is
no longer stored in msmory.
I
Press the SAVE button to display the SETUP pop-up menu.
Verify that the marker before SETUP
3
has disappeared.
10. Generator
»»*
Model 97 only!
*»*
This test checks the built-in generator.
Test equipment:
none
Test setup:
ScopeMMr