Scanning Function
The Single Test versions of the wire map, resistance, HDTDR, and HDTDX
analyzer tests include a scanning function, which you can activate by pressing the
# Scanning ON softkey. The scanning function runs the test repeatedly and
updates the display each time a test is complete. This function is useful for finding
intermittent problems on a cable.
Note
To extend battery life, connect the ac adapter/charger when using
the scanning function for more than 1 minute.
When to Use a Remote Unit
A remote unit is required only when testing twisted pair cable. Table 4-1 shows
which cable tests require a remote and which remotes support each test.
If a remote unit is detected at the start of a Single Test, the test tool runs a wire
map test before running the selected test. If the wire map test fails, the test tool
stops the test and displays the wire map. Press $ Continue Test to run
the selected test.
Note
The DSP-4100 test tool is compatible only with DSP-4100SR units.
The DSP-4100SR unit is not compatible with the DSP-100,
DSP-2000, or DSP-4000 test tools.

Table of Contents

Other manuals for Fluke CableAnalyzer DSP-4000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke CableAnalyzer DSP-4000 Series and is the answer not in the manual?

Fluke CableAnalyzer DSP-4000 Series Specifications

General IconGeneral
BrandFluke
ModelCableAnalyzer DSP-4000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals