Electrical Safety Analyzer
Introduction
XW Warning
To prevent possible electrical shock, fire, or
personal injury, read all “safety information”
before you use the Product.
The Fluke Biomedical ESA615 Electrical Safety Analyzer
(the Product) is a full-featured, compact, portable
analyzer, designed to verify the electrical safety of
medical devices. The Product tests to domestic
(ANSI/AAMI ES1, NFPA 99) and international (IEC62353,
AN/NZS 3551, and parts of IEC 60601-1) electrical-safety
standards. The Product simulates ECG to do
performance tests on ECG monitors.
The integrated ANSI/AAMI ES1 and IEC60601-1 patient
loads are easily selectable.
The Product does these tests:
Line (Mains) voltage
Ground Wire (Protective Earth) resistance
Equipment current
Insulation resistance
Ground (Earth) leakage
Chassis (Enclosure) leakage
Lead to Ground (Patient) and Lead to Lead (Patient
Auxiliary) leakage
Lead isolation (Mains on applied parts leakage)
Differential leakage