Electrical Safety Analyzer
Detailed Specifications
77
Detailed Specifications
Test Standard Selections ........................................ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551
Voltage
Ranges (Mains voltage) ......................................90.0 V to 132.0 V ac rms
180.0 V to 264.0 V ac rms
Range (Point-to-point voltage) ............................0.0 V to 300.0 V ac rms
Accuracy..............................................................±(2 % of reading + 0.2 V)
Earth Resistance
Modes..................................................................2-Wire
Test Current ........................................................>200 mA ac
Range..................................................................0.000 Ω to 2.000 Ω
Accuracy..............................................................±(2 % of reading + 0.015 Ω)
Equipment Current
Range..................................................................0.0 A to 20.0 A ac rms
Accuracy..............................................................±(5 % of reading + (2 counts or 0.2 A, whichever is greater))
Duty cycle............................................................15 A to 20 A, 5 min on/5 min off
10 A to 15 A, 7 min on/3 min off
0 A to 10 A, continuous
Leakage Current
Modes* ................................................................AC+DC (True-rms)
AC only
DC only
* For tests that do not use MAP voltage, AC+DC, AC ONLY, and DC ONLY modes are
available for all leakages. MAP voltages are available only in True-rms (shown as
AC+DC)