Direct Applied Part Leakage Test
Note
The Direct Applied Part Leakage test is available
when the EN62353 standard is selected.
The Direct Applied Part Leakage Current test measures
the leakage current between all applied parts of one
function and the exposed conductive surface on the
housing, to mains earth. For equipment with multiple
applied parts, you must do the test on each group of one
function each in turn while all others float. This test must
only be done for equipment with F-Type applied parts.
For Type B applied part, see direct equipment leakage
schematic in Figure 28.
To do a direct applied part leakage test:
1. Push B.
2. Push the More softkey.
3. Set the desired applied part groupings with G and H.
4. Push the Select softkey. The Direct A.P. test must
already be set.
5. Push E or F to set the applied part test configuration.
6. Push T to apply the test voltage and read the
current in the display.
7. Push E or F to advance to the next group of applied
parts, if applicable.
Figure 29 shows the electrical connections between the
Product and the DUT for a Direct Applied Part Leakage
Current Test.
The outlet conditions below apply when you do this test:
• Normal Polarity
• Reversed Polarity
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1210 Adapter section in this manual.