58
Measure Current
The Product can make dc only, ac only, and ac+dc current
measurements to a maximum of 10 mA. To do a current
measurement:
1. Push the Leakage softkey from the Point-To-Point
menu.
2. Push G or H set the ac only, dc only, or ac+dc
measurement mode.
3. Put test leads in the RED (V/Ω/A ) and BLACK jacks.
4. Put the probe tips on the two points the unknown
current may flow and read the measurement in the
display.
How to Simulate ECG Waveforms
The Product can put different waveforms on the applied
parts connection posts. These signals are used to
measure the performance parameters of ECG monitors
and ECG strip printers. See Figure 35 for the connections
between the Product and an ECG monitor. For monitors
that use the snap style connectors, put the BJ2ECG
adapter into the connectors at the top of the Product and
connect the monitor leads to the snap connectors on the
adapter.
Note
If the ECG monitor/interpreter has banana posts,
use the optional universal snap to banana
adapter to connect to the Product.
To access the ECG Simulation Waveform menu shown in
Figure 34, push C. From this menu, a number of
waveforms are set through F1, and the rate or frequency
of the waveform is set through F2.
gtv109.bmp
Figure 34. ECG Waveform Simulation Menu
To output one of the predefined waveforms, push the
Wave Form softkey. A scroll box with K next to it appears
above the softkey label. Use G or H to scroll through the
different waveforms.