Electrical Safety Analyzer
How to Do Electrical Safety Tests
31
Equipment Current Test
To measure the current consumed by the DUT, push
A. The Product shows the current that flows through
the mains connections of the test receptacle.
Leakage Current Test
The Product measures leakage current for different DUT
configurations. The Product measures leakage found on
the enclosure and the earth connection as well as leakage
on each connected applied part and combinations of
connected applied parts.
The available leakage tests are set by the standard you
set in setup. See the How to Set the Test Standard section
to change the test standard.
Table 5 is a list of six leakage current tests. Their names
are different when the standard is changed in the Product.
Push B to access the leakage current main menu
shown in Figure 19.
Table 5. Test Names Based on Selected Standard
IEC60601 AAMI/NFPA 99
Protective Earth Resistance Ground Wire Resistance
Earth Leakage Current Ground Wire Leakage Current
Touch or Enclosure Leakage Current Chassis Leakage Current
Patient Leakage Current Lead to Ground Leakage Current
Patient Auxiliary Leakage Current Lead to Lead Leakage Current
Mains on Applied Part (MAP) Leakage Current Isolation Leakage Current