ESA615
Users Manual
18
Set the Beeper
To enable or disable the beeper:
1. From the setup menu, push the Instrument Setup
softkey.
2. Push G or H until the Beeper variable is
highlighted.
3. Push e.
4. Push G or H to highlight Off or On.
5. Push e.
Set the Display Contrast
There are two procedures to set the display contrast.
From the Test Sequence start-up menu or through the
setup menu.
When the Product shows its start-up menu (Select a
test…), push G or H to increase or decrease the display
contrast respectively. Push the Done softkey to exit
contrast setup.
To adjust the contrast through the setup menu:
1. From the setup menu, push the Instrument Setup
softkey.
2. Push the Display Contrast softkey.
3. Push G or H to increase or decrease the display
contrast respectively.
4. Push the Done softkey to exit contrast setup.
How to View Instrument Information
To show model number, serial number, firmware version,
and last calibration date of the Product, push Q. Next
push the Instrument Information softkey.
How to Do Electrical Safety Tests
The Product does a number of different electrical and
performance tests on biomedical equipment. The sections
that follow are test descriptions with instructions on how
to do them with the Product.