To do an alternative applied part leakage test:
1. Push B.
2. Push the More softkey.
3. Set the applied part groupings with G and H.
4. Push the Select softkey.
5. Push the Alternative A.P. softkey.
6. Push T to apply the test voltage and read the
current in the display.
7. Push E or F to advance to the next applied part
group(s) of a single function if applicable. Push T
to read leakage current for each group.
Figure 27 shows the electrical connections between the
Product and the DUT for an Alternative Applied Part
Leakage current test.
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1210 Adapter section in this manual.
Direct Equipment Leakage Test
Note
The Direct Equipment Leakage test is available
when the EN62353 standard is selected.
The Direct Equipment Leakage Current test measures the
leakage current between all applied parts and the exposed
conductive surface on the housing to mains earth.
To do a direct equipment test:
1. Push B.
The direct equipment test is the default test and must
already be selected.
2. Push T to apply the voltage and read the leakage
current in the display
Figure 28 shows the electrical connections between the
Product and the DUT for a Direct Equipment Leakage
Current Test.
The outlet conditions below apply when you do this test:
Normal Polarity, Closed Earth
Normal Polarity, Open Earth
Reversed Polarity, Closed Earth
Reversed Polarity, Open Earth
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1210 Adapter section in this manual.