Electrical Safety Analyzer
Instrument Familiarization
9
Table 3. Side and Top-Panel Connections
Item Name Description
1 Equipment Outlet
Equipment outlet, specified to the version of the Product, which supplies a DUT
connection.
2 USB A Controller Port For external keyboard or barcode reader.
3
USB Device Port
(Mini B-style connector)
Digital connection to control the Product from a PC or instrument controller.
4 Fuse Access Door Equipment outlet fuse access.
5 Tilt Stand Holds the Product in a tilted position.
6 SD Card Slot SD Memory Card access
7 AC Power Switch Turns ac power on and off.
8 Power Input Connector
A grounded male three-prong (IEC 60320 C19) connector that accepts the
line-power cord.
9 ECG/Applied Parts Jacks
Connection posts for Device Under Test (DUT) applied parts, such as ECG
leads. Used to test for leakage current through leads and to supply ECG
signals and performance waveforms to a DUT.
10 Banana Jack to ECG Adapter Adapter to connect ECG snap leads to the Product.