Electrical Safety Analyzer
How to Do Electrical Safety Tests
33
The outlet conditions below apply when you do this test:
• Normal Polarity
• Normal Polarity, Open Neutral
• Reversed Polarity
• Reversed Polarity, Open Neutral
IEC60601-1 specifies that the applied parts must be
connected for this measurement. Push E or F to ground
and unground all applied parts connection posts.
Note
A ground connection on an applied part is shown
as a box around the applied part in the display.