Electrical Safety Analyzer
Test Sequences
67
Table 7. Test Settings for Test Sequences (cont.)
Test Setting Description
Default
Value
Multiple PE Tests If set to Yes, the Product prompts you to repeat PE tests or continue. No
Multiple Results
storage
[1]
When set to Store all, all results are put into the test results. If set to Store
worst/last, only the worst measured value is put into the test results. If set to store
worst/last and all measured values are the same, then the last value is put into
the test results.
Store
worst/last
Insulation Test Voltage Sets the insulation test voltage to 250 V dc or 500 V dc. 500 V dc
[1] Applies to leakage current and multiple PE tests only.