Electrical Safety Analyzer
Detailed Specifications
79
Insulation resistance
Ranges................................................................0.5 to 20.0 MΩ
20.0 to 100.0 MΩ
Accuracy
20 MΩ Range..................................................±(2 % of reading + 0.2 MΩ)
100 MΩ Range................................................±(7.5 % of reading + 0.2 MΩ)
Source test voltage .............................................500 or 250 V dc (+20 %, -0 %) 2.0 ±0.25 mA short-circuit current
Maximum load capacitance.................................1 μF
ECG Performance Waveforms
Accuracy
Frequency .......................................................±2 %
Amplitude ........................................................±5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration
Waveforms
ECG Complex .................................................30, 60, 120, 180, and 240 BPM
Ventricular Fibrillation
Square wave (50 % duty cycle).......................0.125 Hz and 2 Hz
Sine wave........................................................10, 40, 50, 60, and 100 Hz
Triangle wave..................................................2 Hz
Pulse (63 ms pulse width) ...............................30 BPM and 60 BPM