Fluke VT650 - Analyzer Connections; Airway Flow (Inlet and Exhaust)

Fluke VT650
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VT650/VT900
Users Manual
12
Analyzer Connections
You can connect the Analyzer to a ventilator and
test lung in either a bi-directional or
unidirectional flow configuration.
Airway Flow (Inlet and Exhaust)
The outer diameter of the airway port is a
standard 22 mm fitting for use with patient
hoses. The inner diameter of the airway port
accepts a 15 mm male respiratory fitting for use
with endotracheal tubes, gas sampling adapters,
and similar equipment.
WCaution
To avoid damage to the Analyzer and
avoid adverse affects on the Analyzer
performance:
Do not put metal objects into
connectors.
Always use the external flow filter on
the main airflow channel inlet. This
helps reduce turbulence and keeps
out small particles that could
damage the flow sensor.
To avoid damage to the sensor,
make sure the pressure inside the
airway port does not exceed 5 psi.
Measure only dry gases with the
airway port. Do not use this port to
measure humidified gases.