Fluke VT650 - Flow Specifications; Airway Flow and Ultra-Low Flow Specs

Fluke VT650
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Gas Flow Analyzer
Specifications
41
Flow
Airway Flow
Resolution...........................0.01<100 slpm and
0.1 >100 slpm
Accuracy .............................see Table 9
Frequency Response..........10 % to 90 % rise time <10 ms
Sample Rate
200 Hz
Dynamic Resistance ..........<2.00 cmH2O @ 60 slpm
Ultra-Low-Flow (VT900)
Range .................................±750 ml/min
Accuracy .............................±1.7 % of rdg or 0.01 slpm
Resolution...........................0.001 lpm
Flow Temperature
Coefficient ...............................Add 0.2 % of reading per °C
between 10 °C and 18 °C
Add 0.2 % of reading per °C
between 28 °C and 40 °C
Table 9. Airway Flow Range and Accuracy
Gas Range Specification
Main Airway Flow
Air, Nitrogen (N2),
Oxygen (O2)
±300 slpm
1.7 % of rdg or
0.04 slpm
Argon, Heliox,
O2 bal He, O2 bal N2
±300 slpm
3.0 % of rdg or
0.08 slpm, typical
Carbon Dioxide (CO2),
Nitrous Oxide (N2O),
O2 bal N20,
±150 slpm
3.0 % of rdg or
0.08 slpm, typical
Ultra-low Flow (VT900)
Air, Nitrogen (N2),
Oxygen (O2)
±0.750 slpm
1.7 % of rdg or
0.01 slpm
Argon ±0.750 slpm
3.0 % of rdg or
0.02 slpm, typical
Nitrous Oxide (N2O) ±0.400 slpm
3.0 % of rdg or
0.02 slpm, typical
Note: Ultra-low flow that is over range can be detected
as unstable. If this occurs, reduce flow.
Table 9. Airway Flow Range and Accuracy (cont.)
Gas Range Specification