Fluke VT650 - Battery and Pressure Specifications; Battery Specifications; Pressure Measurement Specifications

Fluke VT650
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VT650/VT900
Users Manual
40
Battery
Rechargeable Li-Ion
Battery ........................... 10.8 V, 2.5 Ah, 27 Wh, 3ICR19/66
Discharge temperature... 0 °C to 50 °C
Charge temperature ....... 0 °C to 40 °C
Battery life ...................... 8 hours
Battery charge time ........ 5 hours, typical
Note
Battery life is dependent on backlight brightness, auto-
dim, and other battery intensive settings.
Pressure
Ultra Low Pressure (VT900)
Maximum applied
pressure ............................. 400 mbar
Operating Pressure ............ 0 mbar to 10 mbar
Span Accuracy .................. ±1 % of rdg or ±0.01 mbar
(whichever is greater)
Resolution .......................... 0.001 mbar
Low Pressure
Maximum applied
pressure ............................. 400 mbar
Operating Pressure............ (Differential) ±160 mbar
Span Accuracy................... ±0.5 % of rdg or ±0.1 mbar
(whichever is greater)
Resolution .......................... 0.01 mbar
High Pressure
Maximum applied
pressure ............................. 13 bar
Operating Pressure............ -0.8 bar to 10 bar
Span Accuracy................... ±1 % of rdg or ±7 mbar (whichever
is greater)
Resolution .......................... 1 mbar
Airway Pressure
Maximum applied
pressure ............................. 400 mbar
Operating Pressure............ ±160 mbar
Span Accuracy................... ±0.5 % of rdg or ±0.1 mbar
(whichever is greater)
Frequency Response......... 10 % to 90 % rise time <10 ms
Resolution .......................... 0.01 mbar
Sample rate
 200 Hz
Pressure Temperature
Coefficient .............................. Add 0.01 % of range per °C
between 10 °C and 18 °C
Add 0.01 % of range per °C
between 28 °C and 40 °C