VT650/VT900
Users Manual
26
Breath Detect
Sets the options for Breath Detection. Options are:
Mode - Sets the breath detection mode:
Bidirectional
Uni-directional Inspiratory
Uni-directional Expiratory
Off - disables breath triggering
Trigger Source - Sets how the breath detection is triggered:
Flow - Set the Threshold
Pressure - Requires back pressure, such as a test lung
External - enables external TTL triggering input on the Analyzer
(VT900)
Patient - Sets the type of patient:
Adult
Pediatric
Threshold - Enter the numeric value for the detection limit.
+ (Inspiratory threshold)
- (Expiratory threshold)
Table 4. Setup Menu (cont.)