Accuracy by Day of Wear
After the 12 hour start-up period, the sensor can be worn for up to 10 days. To show sensor
performance over time, the absolute relative dierence between the System and reference YSI
glucose values over the wear is presented in Table 9.
Table 9: Dierence Measures by Day (YSI Reference)
Day
Number of CGM-
Reference Pairs
Median Absolute
Relative Dierence (%)
Mean Absolute
Relative Dierence (%)
1
1497 8.7 10.7
4
1470 7.4 9.6
7
1394 7.4 9.1
10
1411 7.5 9.3
The accuracy of CGM versus YSI reference and BG reference was assessed by calculating the
percentage of System readings that were within 15%, 20%, 30% and 40% for reference values
80 mg/dL and above, and 15 mg/dL, 20 mg/dL, 30 mg/dL and 40 mg/dL for values below
80 mg/dL. The results for CGM vs. YSI reference are presented in Table 10.
Table 10: Number and Percent of Results within YSI Reference
Day Number of CGM-
Reference Pairs
Within
±15% /
±15mg/dL
Within
±20% /
±20mg/dL
Within
±30% /
±30mg/dL
Within
±40% /
±40mg/dL
Outside
±40% /
±40mg/dL
1
1497 76.2 87.4 97.9 99.5 0.5
4
1470 82.3 91.4 97.6 99.3 0.7
7
1394 85.0 93.5 98.6 99.3 0.7
10
1411 85.3 92.3 97.9 99.4 0.6
117 118