4-3
Ver.1.36
E-4: Select a mode for interface control. If Manual is selected, the temperature is
controlled at the one set using E-3. If AUTO is selected, the interface temperature is
maintained 100°C above the pyrolyzer furnace temperature; however, the temperature
set by E-3 is the upper limit temperature.
F. Accessories set-up: Sets parameters for accessories such as Selective Sampler, MicroJet
Cryo-Trap, etc. The actual set-ups are dependent upon the analysis mode
selected.
F-1: Sets operating parameters for the accessories to be used.
G. Comments: Comments can be added for each analysis mode.
H. Status bar: Displays the progress of the program.
I. Start/Stop button: Starts or stops the program.
J. Monitoring area: Shows operating condition of each device connected.
J-1: Shows the current temperature of the pyrolyzer furnace.
J-2: Shows the set temperature of pyrolyzer furnace or final temperature set in the
program.
J-3: Shows the current temperature of the interface.
J-4: Shows the set temperature of the interface.
J-5: Shows the temperature zone currently being sampled.
J-6: Show the trap temperature of MicroJet Cryo-Trap.
J-7: Show the elapsed time since the start of program.
J-8: Shows the time required to complete the program.