6-10
Ver.1.36
d The sample cup chuck is a consumable part. Verify that it securely holds the Eco-Stick and that
pressing the release button releases the sample cup. If there is any problem, the sample cup
chuck (PY1-1345) needs to be replaced with a new one. This is customer serviceable, contact
your nearest sales office or Frontier Laboratories.
(4) Sampler base
Contamination or damage of the O-rings (P-12, P/N: PY1-7811) at the sampler base may result in gas
leakage. Clean them by rubbing them with a cotton swab dampened with methanol. Avoid using other
solvents such as acetone, dichloromethane, or etc.
6.2.2 GC injection port
Upon extended use, the injection port split outlet pipe may be clogged by tars. If this occurs, remove it
from the GC and clean it by flushing a solvent through it, and then heat it by a torch while running an
inert gas through it. The pipe needs to be checked every three to four months of use.
6.2.3 Temperature controller
(1) Make sure that connectors for heater power of pyrolyzer, temperature sensor, and temperature
controller have good electrical contact.
(2) Make sure the temperature controller is properly grounded.
6.2.4 When not in use for a long period of time
If the pyrolyzer is not going to be used for a long period of time, remove the pyrolyzer from GC and
store in a nylon bag so that the cartridge heater used in the interface is protected from moisture.
There is a risk of burn injury. Before starting work, make sure that the injector
temperature is below 50°C.
CAUTION
HOT
When cleaning with volatile organic solvents, ensure there is no ignition
source around. When using a burner flame, work in a well-ventilated
area free from flammable organic solvents.
CAUTION
!
Before removing cables, turn off the power switch and unplug the power
code of the temperature controller.
CAUTION
!