6-11
Ver.1.36
6.2.5 List of recommended replacement parts
Pats listed below are consumable parts that are recommended to replace during maintenance work.
The replacement interval raries depending on the usage.
6.2.6 Periodic maintenance and inspection
We recommend that the pyrolyzer be overhauled and inspected every two years and that
hard-to-replace parts such as heater and thermal insulator be replaced for your stable and safety
operation of the product.
Quartz pyrolysis tube
A30
Replacement interval depends on type of
samples, amount used, and number of runs.
Can be heat-cleaned with torch flame.
Replace when pyrolysis tube M30 is replaced.
Replacement recommended when pyrolysis
tube M30 is replaced.
Over tightening injector septum may clog
needle opening with septum debris.
Standard accessory for EGA/PY-3030D
Used when Auto-Shot sampler is used.
For sealing pyrolysis tube M30, 2 pcs used
Replacement not needed when Auto-Shot
Sampler is used all the time.
Septa for liquid
sampler, 20 pcs
Inspection may only be needed if liquid sample
injection with micro syringe is not done.
Used for carrier gas tube joints
Worn-out chuck does not hold eco stick firmly.
Special tool required.
The product needs to be returned to F-Lab.
Replace when trap tube is replaced. 4 pcs
used.
Replace if leakage detected when pyrolyzer
interface or GC injector cools down.
Replace when pyrolyzer is removed from GC
injector.
Other parts that GC
manufacturers
recommend