CHAPTER 2 SPECIFICATIONS
The system consists of the pyrolyzer furnace, which also serves as thermal desorption unit, and
temperature controller.
Multi-Shot pyrolyzer
EGA/PY-3030D
1. Performance guaranteed
(detector: MS)
See CHAPTER 8 of this manual for analytical conditions.
1. Reproducibility of pyrograms
2. Reproducibility of EGA
thermograms
Flash pyrolysis of PS (25 µg) at 550°C (Eco cup LF used), the coefficient of variation
(RSD) of the peak area ratio of styrene trimer (SSS) and methyl stearate (Me-Ste): ≤2%
(Eco cup LF, single cup used)
The coefficient of variation (RSD) of the peak top temperature of PS: ≤0.3%
2. Furnace and sampling unit
(Vertical micro furnace)
Pyrolyzer furnace
Sample introduction
Pyrolysis tube material
Temperature control range and
stability
Cooling system
Cooling time (pressure: 500kPa)
Cylindrical ceramic heater
Push-button to free-fall system with manual up/down-movement
Quartz
Room temperature + 10°C up to 1,050°C (1°C step) / ±0.1°C
Forced air cooling by nitrogen or compressed air
Within 10 min (800°C down to 50°C when ITF in auto mode)
Interface (ITF)
Temperature/Stability
ITF needle
Cartridge heater
40 up to 450°C (1°C step) / ±0.1°C
Needle with side opening (inner surface deactivated by gradient multi-layer treatment)
Sampler
Double-Shot sampler
For Double-Shot and Single-Shot analyses
Liquid sampler
For introduction of liquid samples (regular 10 µL micro syringe can be used)
Sample cup
Eco cup SF (small)
Maximum capacity: 50 µL (deactivated stainless steel)
Eco cup LF (large)
Maximum capacity: 80 µL (deactivated stainless steel)
3. Temperature controller
PC control (PC not included)
Functions
SGL/DBL, Direct EGA, Heart-Cut EGA, ITF temp, Method File, Temp Monitor,
Maintenance
Other functions: control of Selective Sampler, MicroJet Cryo-Trap;, temperature
calibration, and Carrier Gas Selector
Temperature control range
Pyrolyzer furnace
30 up to 1,050°C (1°C step)
Ramp rate
600°C/min max (1°C step)
Hold time
0 up to 999.9 min (0.1 min step)
Pyrolysis time
0 up to 999.9 min (0.1 min step)
Interface
30 up to 450°C (1°C step) Auto/Manual control
Overheat protection temperature
PY: 1100°C, ITF: 500°C, Sample cup standby position: 100°C
Communication
USB 2.0
*: Prolonged operation of the pyrolyzer at a furnace temperature higher than 800°C will shorten the furnace heater life.
When operating at 800°C or higher, keep the operation time less than 20 minutes. If it is operated at 800°C or higher
for a prolonged period time, contact us at cs@frontier-lab.com. When the pyrolyzer is in a standby state, we
recommend that the furnace temperature should be kept below 200°C. This will extend the furnace heater life.