EasyManua.ls Logo

Gatan ORIUS SC 200D - Figure 2-7 Instructions Window; Figure 2-8 Calibration Window

Gatan ORIUS SC 200D
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ORIUS ™ 833 User’s Guide Series Rev 3 2-7
Figure 2-7 Instructions window
A red line will appear on the image.
Position it on a feature of known size.
Press OK on the “Calibrate image” window.
Enter the correct distance for the selected feature (for example 10 line
pairs of cross grating sample where the distance = 10 x 0.463μm) in
the “Calibration” window and select the units (see Figure 2-8).
Figure 2-8 Calibration window
Press OK.
Choose YES in the next window (see Figure 2-9) to complete the
magnification calibration.

Table of Contents