ActionStep
In the System Control module, on the Manual menu, click Execute Manual
Instructions .
1
Result: The Manual instructions dialog box opens.
In the Instructions box, select Alarms:Alarm pre column pressure.2
Click Enabled in the Mode field.3
•
Enter the high pressure limit in the High alarm box.
•
Click Execute.
4
118 ÄKTA avant Operating Instructions 29101556 AB
5 Prepare the system for a run
5.5 Set pressure alarms