Fraction collector tubes and
bottles
The tubes and bottles used in the built-in fraction collector must fulfill the requirements
listed in the following table. Examples of manufacturers are also listed in the table.
Examples of manu-
facturers
Height (mm)Diameter (mm)Tube or bottle
size (ml)
Max.Min.Max.Min.
NUNC™565011.510.53
VWR™76701210.55
BD™ Biosciences,
VWR
1029613.3128
BD Biosciences120114171615
BD Biosciences116110302850
Nalgene™, Kautex™121-L: 64.5
W: 64
1
L: 55
W: 55
1
250 ml bottle
1
Length and width of the rectangular bottle base
Deep well plates, requirements
The deep well plates used in the built-in fraction collector must fulfill the requirements
listed in the table below.
SpecificationProperty
24, 48, or 96No. of wells
Square, not cylindricalShape of wells
10, 5, or 2 mlWell volume
64 ÄKTA avant Operating Instructions 29101556 AB
4 Installation
4.1 Site preparation
4.1.6 Required materials