EasyManua.ls Logo

GE Alpha RT 32551 - Setups, Alignments and Adjustments; Initial Setups

GE Alpha RT 32551
99 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Setups, alignments and adjustments
32551-IMG rev 6 GE Healthcare 47
5 Setups, alignments and adjustments
5.1 SETUP, ALIGNMENT AND ADJUSTMENT PRIOR TO USE
5.1.1 Initial setups
In this section the contents of the nonvolatile memory are checked and
changed if necessary. The setup information is available in two programs,
PAr and SUP. The automatic exposure control information is available in
AEC and tAr programs.
In the SetUP (prog. #52 SUP) the following is setup:
In the PArameter (prog. #53 PAr) the following is setup:
In the Automatic Exposure Control (prog. #64 AEC) information is
available:
In the tArget time (prog. #65 tAr) following information is available:
#52 SUP A Compression autorelease time
#52 SUP L Label type
#52 SUP H Label text
#52 SUP i i/o device
#52 SUP F Default filter
#53 PAr S The serial number of the unit
#53 PAr C Country code
#53 PAr r Tube efficiency in mR/mAs for both filters
#53 PAr b Anode brake selection
#53 PAr t Software mode and Phototimer modes
#53 PAr I Collimator type (Interlock)
#53 PAr d C-arm control board / CD board selection
#64 AEC d Wanted optical density
#64 AEC h IQC - kV programming
#64 AEC I Target time MIN/MAX automatic exposure
#64 AEC c Thickness compensation adjustment
#64 AEC g AEC gain - optical density programming
#64 AEC E Full AEC kV programming (Initial kV vs. breast
thickness)
#64 AEC U kV compensation adjustment
#64 AEC S Density step adjustment
#65 tAr The Alpha RT AA- mode (auto kV) selects the used kV
according to a target time which is programmed for small
and large focus, and for Mo and Rh filters separately. The
target time depends on the compressed breast thickness
and is a linear curve calculated from two programmed
points.

Table of Contents