EasyManua.ls Logo

GE Alpha RT 32551 - AEC Adjustment

GE Alpha RT 32551
99 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Setups, alignments and adjustments
32551-IMG rev 6 GE Healthcare 77
1 Check that the fast turn-around times are equal at both ends. If
adjustment is necessary, move the slotted metal sheet travelling
through the optocoupler.
2 The grid travelling time is adjusted from R5 located on the bucky-
control board. In the CC position travel time of the grid should be 1.6
+/- 0.1 sec. The fast travel time should be 0.17 +/- 0.1 sec.
3 Make sure that the grid moves well in lateral positions (C-arm in +/-
90°angle) as well.
Fig 5.12. Bucky control board
5.2.14 AEC adjustment
Test equipment needed: Sensitometer, Densitometer, Acrylic blocks, Film
& Cassette.
1 Attach Bucky device.
2 Press Den+ & Den- for 3 sec. to enter program (Pr) mode.
CAUTION!
Store all default values in the tables on the next pages (molybdenum filter)
and (rhodium filter).
3 Press Autoexp to AA mode and channel a. Select Molybdenum filter.
4 Select max. kV change range of initial kV (factory default “4 kV) with
kV- and kV+ keys. Value can be selected between 0...9 kV. 0 kV
means that initial kV does not change at all during the exposure.
Value 9 means that kV can decrease or increase 9 kV from initial kV
(which was selected according the breast thickness) during
exposure.
5 Select the internal density correction factor within a range of 50% to
200% (0.50 on screen means 50% and 2.00 on mAs-screen means
200% change to exposure time). The default value is 1.00.

Table of Contents