5 Setups, alignments and adjustments
32551-IMG rev 6 GE Healthcare 53
5.1.4 Service menu program #53 PAr
1 Select the program number 53 (PAr) with kV+ button.
2 Press AUTOEXP to enter the PAr program.
#52SUP/i The density display shows i. In this mode the RS 232 link can
be programmed.
0 = no connections
1 = printer or Alpha ID II connected
2 = PC connected (ONLY WITH Alpha ID [Dataview])
3 = Delta 16 mode
4 = Delta 16 / Alpha ID (Dataview), requires a RS-
switcher board
Press kV+
#52SUP/F The density display shows F. This mode determines which
filter type is selected in the power up. Select the mode using
the density- and density+ keys.
0 = selects automatically the rhodium filter
1 = selects automatically the molybdenum filter
2 = selects the previously used filter
3 = selects the auto-filter mode.
Use mAs- and mAs+ keys to determine the
compression thickness (0 - 95 mm in kV display)
where the filter changes.
Press kV+
Press Autoexp to leave the SUP program and to enter the
special program menu.
#53PAr/S The density display shows S, which means the Serial
Number of the Unit. Check that it equals the serial number
on the label under the image receptor support. If not, correct
using the density- and density+ buttons.
Press kV+