5 Setups, alignments and adjustments
62 GE Healthcare 32551-IMG rev 6
CAUTION!
The maximum and minimum target time values are limited in the #64 AEC
program (parameter I)
8 To store the new values press Den+ and Den- simultaneously until
service mode is entered.
9 Press kV- and Den- until normal mode is entered.
10 Switch OFF the power for a few seconds. Turn the power ON and
ensure that the programming has been stored.
5.1.7 Error and exposure counters (Pr#58 CEC)
The exposure and error counters can be checked in the special program
#58 CEC. The timer display shows the exposure counters and the error
codes: tot, cnt, E01, E02, E03, E04, E07, E08, E09, E10, E11, E12, Slo, Er1,
Er2, Er3, Er4, Er5, Eaf, SYS.
The kV and mAs displays show the number of those errors. Pressing the
mAs- resets the counter in the display.
Pressing the kV+ button selects the next counter. Reset all the counters.
When the timer display shows tot, the exposure counter is shown. It
cannot be reset. The cnt counter is a resettable counter and should be
reset.
Leave the special program by pressing Autoexp.
5.1.8 Adjustment procedure during the installation
NOTE!
REPORTS TO BE RECORDED IN U.S.A. ONLY: Record all requested data
according to the installation report enclosed. Complete the FDA form
2579(3/89) according to the DHR report guide sheet 37139-IMG enclosed.