5 Setups, alignments and adjustments
78 GE Healthcare 32551-IMG rev 6
Density programming will be done later in section 5.2.19, Density
programming.
6 Press Autoexp to select channel b. Repeat instructions in 4 and 5.
7 Press Autoexp to select channel c. Repeat instructions in 4 and 5.
8 Press Autoexp to select semiAuto (A) mode and channel a.
9 Select default kV (factory default 27 kV) with kV+/- keys and the
internal density correction factor (default value 1.00) with mAs- and
mAs+ keys.
10 Press Autoexp to select channel b. Repeat instructions in 10.
11 Press Autoexp to select channel c. Repeat instructions in 10.
12 Press Autoexp to select manual mode. Select default kV (factory
default 27 kV) with kV+/- keys and default mAs (factory default 12
mAs) with mAs- and mAs+ keys.
13 Press Autoexp. Select Rhodium filter and repeat the settings (steps 3
to 13).
Attach the cassette tunnel.
Repeat instructions from 3 to 13 for cassette tunnel.
Attach the magnification tunnel.
Repeat instructions from 3 to 13 for cassette tunnel.