1 Introduction
2 GE Healthcare 32551-IMG rev 6
WARNING!
This X-ray unit may be dangerous to patient and operator unless safe
exposure factors and instructions are observed!
WARNING!
Electric safety: Use fault current protection connector box between unit
power plug and mains. High voltage tank and high voltage connectors are
dangerous areas in the that require special attention as well as 220 VAC
present on Inverter Board, AMD Board and fan connectors.
NOTE!
This unit has been tested at 60 Hz. If the unit will be installed to 50 Hz
mains network, refer to Chapter 5.2.10, Light field intensity adjustment for
collimator light field intensity adjustment and Chapter 4.9, Anode run
voltage check (Pr#57 Ano) or adjustment to 50 Hz mains network for
anode run voltage adjustment.