IRECTION 5141177-100, REVISION 14 BRIGHTSPEED ELITE, EDGE, EXCEL: PRE-INSTALLATION
Page 56 Section 2.0 - Regulatory Clearances
Work Space
Requirement
Minimum Clear
Space
Additional Conditions
Direction of Service
Access (Front of UPS)
914.4 mm (36 in.)* There are no exposed live part hazards with
the cover in place.
This component is typically serviced from
the front with access to the rear.
*If exposed live parts of 151 - 600 volts are
present, 1219 mm (48 in.) is required on
both sides of the workspace with the opera-
tor between.
*If the opposite wall is grounded and
exposed live parts of 151 - 600 volts are
present, 1067 mm (42 in.) is required.
Service Access Width
(Right side and length of
UPS)
762 mm (30 in.) This is the width of the working space in
front of the equipment. A minimum of 762
mm (30 in.) or the width of the equipment,
whichever is greater, is required.
Head Clearance 1981 mm (78 in.) This is the height of the workspace
measured from the floor at the front edge of
the equipment to the ceiling or overhead
obstruction(s). A minimum of 1981 mm
(78 in.) or the height of the equipment,
whichever is greater, is required.
Table 4-7 UPS Subsystem
Work Space
Requirement
Minimum Clear
Space
Additional Conditions
Direction of Service
Access (Front of A1
Disconnect)
914.4 mm (36 in.)* There are no exposed live part hazards with
the cover in place.
This component is typically serviced from
the front with access to the rear.
*If exposed live parts of 151 - 600 volts are
present, 1219 mm (48 in.) is required on
both sides of the workspace with the opera-
tor between.
*If the opposite wall is grounded and
exposed live parts of 151 - 600 volts are
present, 1067 mm (42 in.) is required.
Service Access Width
(Right side and length
of A1 Disconnect)
762 mm (30 in.) This is the width of the working space in
front of the equipment. A minimum of 762
mm (30 in.) or the width of the equipment,
whichever is greater, is required.
Head Clearance 1981 mm (78 in.) This is the height of the workspace
measured from the floor at the front edge of
the equipment to the ceiling or overhead
obstruction(s). A minimum of 1981 mm
(78 in.) or the height of the equipment,
whichever is greater, is required.
Table 4-8 A1 Disconnect Subsystem