GE HEALTHCARE
D
IRECTION 5141177-100, REVISION 14 BRIGHTSPEED ELITE, EDGE, EXCEL: PRE-INSTALLATION
Page 92 Section 10.0 - Radiation Protection
Figure 4-22 Typical Scatter Survey (Head Filter)
0.33
0.33
0.65
0.65
1.3
1.3
2.6
2.6
0.33 0.330.65 0.651.3 1.32.6 2.6
HEAD PHANTOM
Technique:
140 kV
100mA
1 Sec
4 x 5.00mm
Technique:
140 kV
100mA
1 Sec
4 x 5.00mm
50cm
50cm
50cm
50cm
HEAD PHANTOM
ISO–Contour Levels: 0.33, 0.65, 1.3 and 2.6 µGy/Scan
ISO–Contour Levels: 0.33, 0.65, 1.3 and 2.6 µGy/Scan