EasyManuals Logo

GE Dash 4000 Operator's Manual

GE Dash 4000
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #218 background imageLoading...
Page #218 background image
C-4 Dash™ 3000/4000/5000 2000966-386D
Factory Defaults
Parameter Limits
The following are the parameter alarm limit factory defaults. To access these options,
select MORE MENUS > MONITOR SETUP > MONITOR DEFAULTS > SETUP
DEFAULT LIMITS.
ADULT-ICU NEONATAL-ICU OPERATING ROOM
LOW HIGH LOW HIGH LOW HIGH
HR 50 150 90 200 –1 150
PVC/MIN 6 6 6
ST-I
–2.0 2.0 –2.0 2.0 –2.0 2.0
ST-II
ST-III
ST-V1
ST-AVL
ST-AVF
ST-AVR
ST-V2
ST-V3
ST-V4
ST-V5
ST-V6
NBP-S 80 200 40 100 40 200
NBP-D 20 120 20 60 20 120
NBP-M 40 140 30 70 40 140
ART-S 80 200 40 100 40 200
ART-D 20 120 20 60 20 120
ART-M 40 140 30 70 40 140
ART-R 50 150 90 200 –1 150
FEM-S 80 200 40 100 40 200
FEM-D 20 120 20 60 20 120
FEM-M 40 140 30 70 40 140
FEM-R 50 150 90 200 –1 150
UAC-S 80 200 40 100 40 200
UAC-D 20 120 20 60 20 120
UAC-M 40 140 30 70 40 140

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable