Factory Defaults
2000966-386D Dash™ 3000/4000/5000 C-5
UAC-R 50 150 90 200 –1 150
PA-S
–99 350 –99 350 –99 350
PA-D
PA-M
CVP
RA
UVC
LA
ICP
SP
CO2-EXP 150–150–150
CO2-INSP 5—5—5
CO2-RESP 5 30 15 100 1 200
NO BREATH 20 15 30
SPO2 90 105 88 100 90 105
SPO2-R 50 150 90 200 –1 150
BT 30.0°C/
86.0°F
42.0°C/
107.6°F
30.0°C/
86.0°F
42.0°C/
107.6°F
0.0°C/
32.0°F
42.C/
107.6°F
SVO2 60 80 60 80 60 80
RR 5 30 15 100 1200
RR-APNEA 20 15 20
TEMP 1 30.0°C/
86.0°F
42.0°C/
107.6°F
30.0°C/
86.0°F
42.0°C/
107.6°F
0.0°C/
32.0°F
42.C/
107.6°F
TEMP 2 30.0°C/
86.0°F
42.0°C/
107.6°F
30.0°C/
86.0°F
42.0°C/
107.6°F
0.0°C/
32.0°F
42.C/
107.6°F
O2-INSP 18 102 18 102 18 102
O2-EXP –1 102 –1 102 –1 102
ADULT-ICU NEONATAL-ICU OPERATING ROOM
LOW HIGH LOW HIGH LOW HIGH

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable