Programming Site Data 1-53
March 2008
Option Card RTD Inputs Option cards with RTD inputs have a temperature range of –100° to
350°C. Complete the following steps to set up the two RTD inputs of
an option card installed in Slot x:
1. Press
[F1] to set up RTD input A or [F2] to set up RTD input B.
Note: The set up of RTD input A is used as an example. Identical
procedures would be used to set up RTD input B.
2. Enter a Label of up to eight characters for input A and press [ENT].
3. At the Input Measurement prompt, press
[F1]=OFF to disable RTD
input A and return to the previous prompt, or press
[F2]=TEMP to
enable RTD input A.
4. Enter a value for the Base (low) end of the chosen input range and
press
[ENT].
5. Enter a value for the Full Scale (high) end of the chosen input
range and press
[ENT].
6. Press
[F1]-[F2] to set up another input or press [EXIT] to select
another slot for set up.
Procedure Options After completing the above steps, the meter returns to the Global
PROGRAM
. Continue as follows:
• To continue programming the meter, refer to the menu maps in
Appendix A and navigate to the desired menu. Then, proceed to
the appropriate section of this manual for instructions.
• To leave the User PROGRAM and retain the previous settings, press
[EXIT] twice and then press [F1] = NO at the SAVE prompt. Any
programming changes will be discarded and you will be returned
to the data display
.
• To leave the User PROGRAM and return to measurement mode, press
[EXIT] twice and then press [F2] = YES at the SAVE prompt. Your
programming changes will be entered into the meter’s memory,
and you will be returned to the data display.