EasyManua.ls Logo

GE Druck MCX-II - Special Measurement Functions; Scaled Readings; Setting up Scaled Readings (Linear Relationship)

GE Druck MCX-II
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
70
K0320 Issue No. 1
Scaled Readings
Except in temperature measure ranges
all readings can be presented in a
number of 5 digits and a sign.
This number could represent an
engineering unit like gallons/hour,
revolutions/minute or any other
relationship. This facility is available
on both measurement and source
functions. The original signal reading
is still displayed in smaller sized digits.
Setting up scaled readings
(linear relationship)
Select SCALE to open the set up
screen. Fill out the high- and low-limits
for the span (preset at 4 and 20 mA in
the milliamp ranges). Fill out both
scale ends for the new scale. Press
EXE to show the working screen.
NOTE:
Scaling in lower ranges only
when auto ranging.
Special Measurement Functions

Table of Contents

Related product manuals