102
K0320 Issue No. 1
Transmitter calibration
To check or calibrate a transmitter
All normal key-pad functions are
available to change the output. Use
the
key to change to fixed steps.
Use and keys to step through
0%, 25%, 50%, 75% and 100% of
the span as set-up. Press
to
return to the normal key-pad functions.
An ideal transmitter will show
equal readings. The difference of
both readings is expressed in % of
the transmitter span or reading and
is shown in the upper window in
small size digits.