80
K0320 Issue No. 1
Transmitter (XMTR) Simulation
Simulieren 2-Leiter Transmitter
Simuler un transmitter 2-fils
Volts output
Volt Geben
Sortie volt
Volts
Go to Volts in the DIRECT mode to
output any signal between zero and
12 Volts. Adjustments are made with
100 microvolts. If the adjusted output
level cannot be maintained the screen
will prompt the warning CHECK LOOP.
Milliamps/XMTR
Use mA terminals to source a current
into a resistor (active mode). Use
XMTR terminals to simulate a 2-wire
transmitter (passive mode). Go to
milliamps in the DIRECT mode to
output any signal between 0 and
24 mA. Adjustments are made with
1 microamps resolution. If the
adjusted output level cannot be
maintained the screen will prompt the
warning CHECK LOOP. To output
fixed 0, 4, 8, 12, 16 and 20 mA levels,
press
key to change to fixed steps.
Use
and keys to adjust the output
current. Press
key to return to
normal output mode.
Output of electrical signals