78
K0320 Issue No. 1
Millivolts
Go to millivolts in the DIRECT mode to
output any signal between -10 mV and
100 mV. Adjustments are made with
1 microvolt resolution.
If the adjusted output level cannot be
maintained the screen will prompt the
warning CHECK LOOP.
NOTE:
‘Terminal to test lead’
junctions may create e.m.f.
microvolts causing an offset in the
actual output signal.
Millivolts output
Millivolt Geben
Sortie millivolts
Output of electrical signals