GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 4-3 - General Procedure 4-21
4-3-11-2 Backup System Presets and Configurations
NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need 
to be reloaded, after the software update, they will be the same ones the customer was using prior to 
service.
1.) Insert a formatted CD-R or DVD-R into the drive.
2.) Press UTILITY
 key 
3.) Select SYSTEM
. on the monitor screen
4.) On the monitor display, select BACKUP/RESTORE
.
5.) In the Backup list, select Patient Archive, Report Archive, User Defined Configuration.
6.) In the Media field, select DVD-R.
7.) Select BACKUP
.
The system performs the backup. As it proceeds, status information is displayed on the Backup/Restore 
screen.
Figure 4-16   Backup/Restore Menu
Check here to 
backup presets 
and configurations