GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 4-9 - Board Function Checks 4-47
4-9-4 SYSCONML Assy Function Validation Procedure
4-9-4-1 L1SYSCON diagnostic Test
1.) Turn on the system and check the completion of echoloader loading.
2.) Do the Service Login. Refer to 5-10-2-2 "Service Login" on page 5-41.
3.) Select 'Diagnostics' => 'LOGIQ P5 Diagnostics' => 'FRU test' => 'L1SYSCON' => 'L1SYSCON
Overall'.
4.) Execute the diagnostic test.
5.) All diagnostic test item of L1SYSCON should be passed.
6.) Shut down the system.
7.) Restart the system
8.) Check the basic function of system's different modes. Refer to 4-3-4 "System B/M-Mode Checks"
on page 4-8 and refer to 4-3-5 "System CFM and PWD Checks" on page 4-9.
4-9-4-2 Peripheral Test
Refer to Section 4-5 "Peripheral Checks" on page 4-26.
Figure 4-46 L2SYSCON Diagnostic Test