GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 5-10 - Service Platform 5-43
5-10-4 Error Logs Tab
From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit 
selection. The Service Interface allows scanner logs to be viewed by all service users.
The 
Filter Error log is keyed to GE Service login access permissions and is not available to customer 
level analysis.
The log entries are color-coded to identify the error level severity at a glance.
The Service Interface supports the transfer of these logs to local destinations such as the MOD, and 
CD-ROM drive. 
Table 5-6    Log Entry Key
Severity Error Level Color Code
1
Information
Green
2
Warning
Blue
3
Error
Red
Figure 5-41   Log Viewer / Logs / Log Entries