8-76 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-30 P3RLY / P2RLY Assy & P2RLY with Dummy
8-2-30-1 Tools
• Common pilIips screwdrivers
8-2-30-2 Preparations
• Shut down the system and switch off the main breaker.
8-2-30-3 Removal procedure
1.) Remove the side right cover. Refer to the 8-2-16 "Side Right Cover" on page 8-52.
2.) Remove the side left cover. Refer to the 8-2-15 "Side Left Cover" on page 8-51.
3.) Remove the Front Cover. Refer to the 8-2-17 "Front Cover" on page 8-53.
4.) Remove the ASIG Board. Refer to the 8-2-35 "ASIG Assy" on page 8-86.
5.) Unscrew 8 screws (1-8) and remove the metal cover.
6.) Unscrew 6 screws (1-6) and remove the PCB.
Figure 8-111 RLY Cover
Figure 8-112 Screws of the RLY board