8-184 Section 8-7 - Mechanical Option Installation instruction
8-7-5 Cable Arm Hook installation
1.) Cable Arm Hook should be installed in prior to installing Probe Cable Hanger, or Endo Probe 
Holder.
2.) Unscrew 1 existing screw(2159634, BH M4x10 WHT) from the rear side of the keyboard. 
3.) Screw 2 screws(2306565 BH M4x16 WHT, 2159634 BH M4x10 WHT) to fix the Cable Arm hook. 
User longer screw provide on the upper hole, and use existing screw removed in the step 2 on the 
figure below. 
Figure 8-256   Cable Arm Hook installation
Figure 8-257   Cable Arm Hook installation